If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual TEKTRONIX PASSIVE HIGH-VOLTAGE PROBES P5122. We hope that this TEKTRONIX PASSIVE HIGH-VOLTAGE PROBES P5122 user guide will be useful to you.
Manual abstract: user guide TEKTRONIX PASSIVE HIGH-VOLTAGE PROBES P5122DATA SHEET
Detailed instructions for use are in the User's Guide.
[. . . ] Passive High-voltage Probes
P5100 · P5120 · P5122 · P6015A Data Sheet
P5122
For TPS2000 Series Only DC to 200 MHz 1000 VRMS CATII when DC-coupled*1 Floatable up to 600 VRMS CATIII 100X IEC61010-031
P6015A
High Voltage 20 kV DC / 40 kV Peak (100 ms Pulse Width) High Bandwidth 75 MHz
P5120
Silicone Dielectric Optional 1000X Readout Coding Wide Compensation Range (7-49 pF) Heavy-duty Versatile Ground Lead and Alligator Clip
Features & Benefits
P5100
DC to 250 MHz 1000 V CAT III (2500 V DC + Peak AC) 100X with Readout Coding UL3111-1, EN61010-1, IEC61010-2-031, CSA1010. 1, CSA1010. 2. 031 7 to 30 pF Compensation Range
Applications
Power Supply Design Design Motor Drive Electronic Ballast Power Semiconductors Switch Mode Control UPS Systems Power Converters
*1 The P5122 probe should not be used for AC-coupled measurements on signals with greater than 300 V DC offset. [. . . ] The probe can be compensated to match plug-ins and oscilloscopes with nominal input capacitances of 7-30 pF. A variety of screw-on accessories provide easy connection to the device under test.
The P6015A uses an environmentally safe silicone compound for a dielectric and never needs refilling. Other features include: A 7-49 pF compensation range, small compensation box that fits on adjacent amplifier inputs, and a readout option for use with most Tektronix digital scopes. With the readout option, displayed voltage amplitude values will be the actual signal value rather than understated by a factor of 1, 000. Note: Using the readout version with other than Tektronix digital scopes may result in an erroneous readout display.
2
www. tektronix. com
Passive High-voltage Probes -- P5100 · P5120 · P5122 · P6015A
Characteristics
P5100/P5120/P5122/P6015A
Probe Nominal Length Attenuation Bandwidth Rise Time (Typical) Loading Max. 10 ft.
100X 20X 100X 1000X 1000X
250 MHz 200 MHz 200 MHz 75 MHz 75 MHz
1. 75 ns 2. 2 ns 2. 2 ns 4. 0 ns 4. 0 ns
10 M / 2. 75 pF 5 M / 11. 2 pF 100 M / 4. 0 pF 100 M / 3 pF 100 M / 3 pF
2. 5 kV N/A (DC + Pk AC) 1000 V CAT III 1000 VRMS 600 VRMSCAT II, CATII 300 VRMSCAT III 1000 VRMS 600 VRMS CATIII CATII 20 kV N/A 20 kV N/A
7 to 30 15 to 25 10 to 22 7 to 49 7 to 49
Yes No No No Yes
Ordering Information
P5100
100X, 10 ft. ground lead, instruction manual (070-8151-04).
Includes: Cable markers set, crocodile clip, adjustment tool, hooktip (small), hooktip
Recommended Accessories for P5100, P5120, P5122:
See P5100, P5120, P5122 Probe Accessories.
P6015A
Recommended Accessories
Probe Tip to BNC Adapter Order 013-0291-xx. Includes: Hook probe tip (206-0463-xx); banana plug tip (134-0016-xx); crocodile clip plugs onto ground lead (344-0461-xx); ground lead (196-3363-xx); carrying case (016-1147-xx); instruction manual (070-8223-xx). [. . . ] All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies. [. . . ]